(b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Event related to mw # 2210968-2022-09939.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Product complaint # (b)(4).Date sent to the fda: 1/20/2023.H6 component code: g07002 no device problem found.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that one empty winding former, one paper lid, one needle-suture piece, and one suture piece that pertains to product code d8350 were received for evaluation.In order to evaluate the conditions of the returned sample, the suture pieces were observed with the ends cut.Body fluids and crimping marks were observed on the surface of the suture that was possibly caused by a surgical instrument.The condition of the samples received indicates improper handling of the device and no functional test was performed.As with any device, care should be taken to avoid damage to the strand when handling.Avoid the crushing or crimping action of surgical instruments, such as needle holders and forceps.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Additional h3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that a closed packet that pertains to product code d8350 was received for evaluation.Upon initial inspection, of the sample, no external damages were observed on the packet.Visual inspection of the unopened sample, no defects were found on the packet.The sample was opened, and the swage and attachment area were noted to be as expected.The suture was dispensed without problems and examined along the strand and no defects, damage, or suture breakage were noted.A functional test was performed using an instron equipment and the tensile strength force was above the minimum requirement.The event described could not be confirmed as the device performed without any defect noted.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|