Model Number 71420552
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Loss of Range of Motion (2032); Fibrosis (3167)
Date of Event 11/07/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference case(b)(4).
 
Event or Problem Description
It was reported that, after a tkr, a patient presented with fibrosis causing pain and reduced range of motion (knee).This adverse event was treated by a revision surgery on (b)(6) 2022, in which a gns ii con ins sz7-8 11mm was removed and patient´s knee was debrided of fibrosis tissues.A new tibial insert was placed.Patient's current health status is unknown.
 
Additional Manufacturer Narrative
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, per complaint details, an insert revision/exchange was performed due to fibrosis causing pain and reduced range of motion.Reportedly, the surgeon does not fault the device and is happy with the surgical outcome.It was communicated that the requested medical documentation is not available.Without the requested clinical documentation, definitive clinical factors could not be concluded to have contributed to the event; however, fibrosis is a known post-surgical occurrence, and it cannot be concluded that there was a malperformance of the implant(s).The patient's current health status is unknown.The patient impact beyond the reported fibrosis, pain, and reduced range of motion with subsequent debridement and insert revision cannot be determined.No further medical assessment can be rendered at this time.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for knee systems revealed that decreased range of motion as a result of trauma, improper implant selection, improper implant positioning, improper fixation, and/or migration of the components, this has been identified as possible adverse effects.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors that could contribute to the reported event include post-operative healing issue and/or patient condition.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GNS II CON INS SZ7-8 11MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15901709
Report Number1020279-2022-04882
Device Sequence Number5016327
Product Code JWH
UDI-Device Identifier03596010208187
UDI-Public03596010208187
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K951987
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2025
Device Model Number71420552
Device Catalogue Number71420552
Device Lot Number15GM10148
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/07/2022
Supplement Date Received by Manufacturer12/21/2022
Initial Report FDA Received Date12/02/2022
Supplement Report FDA Received Date12/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/17/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-