• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD Q-SYTE¿ VIAL ACCESS ADAPTER; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD Q-SYTE¿ VIAL ACCESS ADAPTER; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number 385108
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/03/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that yellowish foreign matter was found inside the middle chamber of the bd q-syte¿ vial access adapter.The following information was provided by the initial reporter: "it has been noted to have a yellowish substance in one of the chambers ( middle chamber ) in this product.It was reported by a co-worker.".
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval? yes.D10: returned to manufacturer on: (b)(6) 2023.Investigation summary : our quality engineer inspected the 5 samples submitted for evaluation.The reported issue of foreign matter was confirmed upon inspection of the samples.Analysis of the samples showed that there was excessive adhesive on the samples.This catalog is glued together in the union between the vial adapter and the q-syte assembly.What is being observed is uncured adhesive, which has normal color of a yellow tone.Sometimes when there is an excessive amount of adhesive it may not cure completely causing the yellowish tone.The excess adhesive does not affect the functionality of the device.Production records were reviewed, and this batch was in compliance with our product specification requirements.
 
Event or Problem Description
It was reported that yellowish foreign matter was found inside the middle chamber of the bd q-syte¿ vial access adapter.The following information was provided by the initial reporter: "it has been noted to have a yellowish substance in one of the chambers ( middle chamber ) in this product.It was reported by a co-worker.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD Q-SYTE¿ VIAL ACCESS ADAPTER
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX 
Manufacturer (Section G)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX  
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15902251
Report Number9610847-2022-00448
Device Sequence Number17419073
Product Code FPA
UDI-Device Identifier00382903851089
UDI-Public(01)00382903851089
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K013621
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date02/28/2026
Device Catalogue Number385108
Device Lot Number1053411
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/03/2022
Supplement Date Received by Manufacturer02/03/2023
Initial Report FDA Received Date12/02/2022
Supplement Report FDA Received Date02/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-