• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 REFURB, SHERLOCK 3CG TCS; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 REFURB, SHERLOCK 3CG TCS; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number 9770131R
Medical Device Problem Codes Failure to Calibrate (2440); Data Problem (3196)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/23/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The device has not been received by the manufacturer for evaluation.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event or Problem Description
Sensor not picking up the 3cg rhythm and the tip.
 
Event or Problem Description
Sensor not picking up the 3cg rhythm and the tip.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of sensor not picking up the 3cg rhythm and the tip was confirmed.The reported issue that the sensor is not picking up 3cg rhythm and the tip is confirmed.The sensor is not recognized when connected to a display, causing the ecg and magnet tracking to not function.Reported issue root cause: the root cause is due to an internal failure within the sensor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REFURB, SHERLOCK 3CG TCS
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer Contact
becky garcia
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key15902298
Report Number3006260740-2022-05561
Device Sequence Number20112091
Product Code LJS
UDI-Device Identifier00801741125300
UDI-Public(01)00801741125300
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K113808
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9770131R
Device Catalogue Number9770131R
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/23/2022
Supplement Date Received by Manufacturer01/24/2023
Initial Report FDA Received Date12/02/2022
Supplement Report FDA Received Date01/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-