EXACTECH, INC. LOGIC CR FEMORAL CEM, LEFT, SZ 4; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
|
Back to Search Results |
|
| Model Number |
LOGIC CR FEMORAL CEM, LEFT, SZ 4 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Insufficient Information (4580)
|
| Date of Event |
11/15/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported, approximately 6 years post initial left side tka, the male patient presented for a revision knee due to femoral loosening.Femur debonded from cement during extraction.Patient received full revision of femur component.Patient was last known to be in stable condition following the event.Images received.The device is not available for evaluation due to hospital policy.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
After further review of additional information received the following sections g3, g6, h2, h3 and h6 have been updated accordingly.(h3) the revision due to aseptic femoral loosening reported most likely was the result of mechanical loss of fixation.Contributions from cement technique, excessive resection of the femoral bone and/or patient conditions to the failure cannot be determined from the provided information and the devices were not available for evaluation.The most probable root cause associated with the reported event of ¿loosening - femoral¿ is associated with weakened integration of the device at the bone-implant interface due to loss of fibrous and/or bony tissue and leading to compromised anchorage of the device.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: d8, d10, e4 (remove erroneous entry from initial report), h6 clinical code, component code, types of investigation findings and investigation conclusions.D10:
lgc tibial fit tray cem sz 4f / 4t, (category#: 02-012-45-4040, serial#: (b)(6),
logic cr tib insert std, sz 4, 11mm, (category#: 02-012-47-4011, serial#: (b)(6),
three peg patella 38mm, (category#: 200-02-38, serial#: (b)(6).If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|