As the lot number for the device was provided, a review of the device history record is currently being performed.The device has not been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.Expiry date: 11/2023.
|
|
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.Three photos were provided and reviewed.The first and third photos show a portion of the catheter completely detached with the balloon still inserted in the introducer sheath.A hemostat is seen clamped near the detached portion of the catheter shaft.The second photo shows the health professional holding the sheath which appears inserted in the patient.The detached catheter segment with the balloon inserted in the sheath can be seen.In the other hand, the health professional can be seen holding the other detached segment of the catheter with the guidewire still inserted in it.Therefore, the investigation is confirmed for the reported sheath removal difficulty and identified catheter detachment as the catheter is seen completely detached with the balloon still inserted in the introducer sheath from the provided phots.However, the investigation is inconclusive for the reported sheath insertion difficulty as no evidence of the failure could be determined.A definitive root cause for the reported sheath insertion difficulty, sheath removal difficulty and identified catheter detachment could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: b5, d4 (expiry date: 11/2023), g3, h6 (device, method).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
It was reported that during an angioplasty procedure in superficial femoral artery, the pta balloon dilatation catheter was allegedly difficult to insert and had incompatibility issues.It was further reported that the balloon cannot get out from the sheath.The physician removed the balloon, sheath and guidewire all out.There was no reported patient injury.
|