• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN ABRE VENOUS SELF-EXPANDING STENT SYSTEM; STENT, ILIAC VEIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN ABRE VENOUS SELF-EXPANDING STENT SYSTEM; STENT, ILIAC VEIN Back to Search Results
Model Number AB9U20080090
Medical Device Problem Code Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/29/2022
Type of Reportable Event Malfunction
Event or Problem Description
A physician was implanting an abre self-expanding stent for the treatment of a soft tissue lesion with (b)(4)% stenosis in the left mid/ proximal region of the common external iliac vein.There was mild tortuosity.A non-medtronic 9fr sheath was used.The device was prepped as per the ifu with no issues identified.The lock-pin was removed after advancing to the lesion prior to deployment.It was reported the proximal 15 mm of the stent did not initially expand so the delivery system could not be removed from the patient for a few minutes until the stent popped open.The sheath was slightly kinked (non- medtronic sheath).The physician also did not pre-dilate. due to nitinol material and perhaps radial strength, the top of the stent eventually popped open.This was after approximately 5 minutes of wiggling around the delivery system/advancing it forward and backwards gently.Once stent fully expanded and delivery catheter was removed with difficulty from the patient and post dilated with an 18x60 atlas balloon.No deformation was noted to the deployed stent.Confirmed stent placement with ivus and closed with manual pressure.There was no vessel damage.No patient injury was reported.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Still image evaluation: image were provided for evaluation.The top of the stent can be observed to not have fully deployed.Kinks were observed to the non-medtronic sheath the stent is fully deployed and no damage was observed to the abre delivery system.The reported event can be confirmed from the images attached for evaluation.Product analysis a kink was observed on the inner tubing the stent was deployed.Approximately 80mm of the inner tubing was exposed medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABRE VENOUS SELF-EXPANDING STENT SYSTEM
Common Device Name
STENT, ILIAC VEIN
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key15903242
Report Number2183870-2022-00412
Device Sequence Number5427571
Product Code QAN
UDI-Device Identifier00643169796447
UDI-Public00643169796447
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P200026
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/24/2025
Device Model NumberAB9U20080090
Device Catalogue NumberAB9U20080090
Device Lot NumberB436891
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/29/2022
Supplement Date Received by Manufacturer01/18/2023
Initial Report FDA Received Date12/02/2022
Supplement Report FDA Received Date01/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/25/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-