Additional information has been received for this event and for the evaluation.This supplemental report is being submitted to provide this information.Information on devices used in conjunction with the device for this event is not known.Device history record review indicates that the device was manufactured and tested in accordance with all applicable procedures and met all final product release criteria at inspection.The exact cause of the event could not be exclusively identified.However based on the device evaluation and the past investigation results, it is likely the probe break possibly occurred as below: grasping section was closed without grasping anything while the output in seal and cut mode was activated, (this includes after tissue resection) causing the tissue pad to wear and tear away.The non-insulated area of the grasping section and the probe came into contact due to wear of the tissue pad.The output in seal and cut was activated while the non-insulated area of the grasping section was in contact with the probe.As a result, a contact mark was developed.A force to activate the output in seal and cut mode, or a force to grasp tissue was applied to the probe.Therefore, the cracks developed at a contact mark.A force was applied to the probe causing it to break.The coating of the grasping section could have come off when dirt such as burn was scraped off with something hard.The instructions for use includes the following statements that warn against the issue: do not activate output in seal and cut mode while the grasping section is closed without contacting tissue or vessel, or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.When cutting and vessel sealing is performed in seal and cut mode, apply light tension on the tissue so that users can confirm it is transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall off, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation.If the grasping section, metal-exposed area around it or the probe tip gets tissue stuck during treatment, wipe it with a soft object such as a piece of gauze or a brush.Do not attempt to scrape it with a sharp object such as a scalpel or the tip of tweezers.Otherwise, the grasping section, metal-exposed area around it, the fluorine resin part, a coated surface or the probe tip may be scratched and damaged, which may lead to fall-off of the damaged part into the body cavity or burns of the tissue by a high-frequency leak current output due to destruction of the insulation structure.
|