Additional information has been received for this event.The device has been evaluated.This supplemental report is being submitted to provide this information.The device evaluation has been completed.The user¿s complaint of broken probe was not confirmed.Upon inspection and testing, it was observed that the tissue pad in the grasping section was worn away, and a part of the tissue pad was peeled away.The tissue pad in the grasping section was torn with no missing part.There was no crack and breakage in the probe.The probe was scratched with peeling coating and presence of contact mark.When the probe check was performed, no abnormality occurred, and the device could output.The grasping section jaw had coating peeling and scratched.
|
|
Addendum dec 5, 2022: the device probe broke off in the patient.The broken piece was successfully retrieved from the patient body laparoscopically.The procedure was completed with approximately five minutes delay with no clinical relevance with a new similar device.Addendum dec 15, 2022: upon evaluation of the device, it was observed that the tissue pad in the grasping section was worn and peeled back.There was no missing part to the tissue pad.This is a reportable malfunction of the device.
|
|
Additional information has been received for this event and for the evaluation.This supplemental report is being submitted to provide this information.Please see the updates in sections: b3, b5, g2, g3, g6, h2, h6, and h10.Information on devices used in conjunction with the device for this event is not known.Device history record review indicates that the device was manufactured and tested in accordance with all applicable procedures and met all final product release criteria at inpsection.Since the probe broken was confirmed with a similar complaint, (b)(4), we assume that this complaint is the second product.The exact cause of the event could not be exclusively identified.However based on the device evaluation and the past investigation results, it is likely the peeling of the tissue pad occurred by the following reason: grasping section was closed without grasping anything while the output in seal and cut mode was activated, (this includes after tissue resection) causing the tissue pad to wear and peel off.The coating of the grasping section could have come off when dirt such as burn was scraped off with something hard.It has been confirmed that the probe coating peels off if the device was handled as follows.However, it could not be conclusively determined if such handling actually caused the reported phenomenon.Activate the device in seal and cut mode for a long duration while no tissue is grasped by the grasping section and the distal end of the probe.Activation in seal and cut mode continued after completion of a tissue cutting is also included in such activation.The instructions for use includes the following statements that warn against the issue: ·do not activate output in seal and cut mode while the grasping section is closed without contacting tissue or vessel, or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.When cutting and vessel sealing is performed in seal and cut mode, apply light tension on the tissue so that users can confirm it is transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall off, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation.If the grasping section, metal-exposed area around it or the probe tip gets tissue stuck during treatment, wipe it with a soft object such as a piece of gauze or a brush.Do not attempt to scrape it with a sharp object such as a scalpel or the tip of tweezers.Otherwise, the grasping section, metal-exposed area around it, the fluorine resin part, a coated surface or the probe tip may be scratched and damaged, which may lead to fall-off of the damaged part into the body cavity or burns of the tissue by a high-frequency leak current output due to destruction of the insulation structure.
|