Model Number N/A
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 11/02/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device was discarded by the facility.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Additional associated products with mdrs: 42532006702, tibia cemented 5 degree stemmed rt size d, lot# 64682462.Mdr: 3007963827-2022-00297.Associated products: 42522100310, articular surface medial congruent (mc) right 10mm, lot# 64323804; 42540000032, all poly patella cemented 32 mm diameter; 42509902525, 2.5 mm female hex screw 25 mm length, lot# 64880177; 00579104100, headed screw 48 mm length, lot# 64874198; 00590102000, headless trocar drill pin 3.2 mm diameter 75 mm length, lot# 64809345.
 
Event or Problem Description
It was reported the patient underwent right tka procedure.Subsequently the patient was revised for femoral and tibial loosening approximately 2 years post implantation.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.  h6-component code- suggested code: mechanical (g04) - femur.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no related deviations or anomalies during manufacturing.No compatibility issues were noted.Review of complaint history found no additional related issues for this item and the reported part and lot combination.X-ray evaluation by third party hcp states it is difficult to confirm or exclude a smaller periprosthetic lucency in the femoral component.This complaint cannot be confirmed.Root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMUR CEMENTED CRUCIATE RETAINING (CR) STD RIGHT SIZE 5
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15904063
Report Number3007963827-2022-00296
Device Sequence Number5914578
Product Code JWH
UDI-Device Identifier00889024230323
UDI-Public(01)00889024230323(17)300228(10)64586991
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K113369
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42502605802
Device Lot Number64586991
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/04/2022
Supplement Date Received by Manufacturer02/03/2023
Initial Report FDA Received Date12/02/2022
Supplement Report FDA Received Date02/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/21/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexFemale
-
-