| Model Number |
C503 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The last calibration performed on 21-oct-2022 was acceptable.Quality controls were within specified ranges.Instrument performance testing was completed with results outside of specification.The field service engineer checked the instrument and determined a small crust of serum on the sample probe.He cleaned and adjusted the sample probe.Instrument performance testing was run, which did still not pass within specifications.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
The initial reporter stated they received discrepant results for one urine patient sample tested with tpuc3 (total protein urine/csf gen.3) on a cobas c 503 analytical unit.The sample initially resulted in a tpuc3 value of 0.07 g/l.This result was reported outside of the laboratory and questioned by the physician.The result is inconsistent with the microalbumin result of 273 mg/l of the patient.The sample was repeated on the next day on a second analyzer resulting in a tpuc3 value of 0.35 g/l.This result is consistent with the microalbumin result of 274 mg/l of the patient.The tpuc3 reagent lot was 627560 with an expiration date of 31-aug-2023.
|
| |
|
Additional Manufacturer Narrative
|
|
The field service engineer adjusted the naoh level and adjusted the ultrasonic mixer.Instrument performance testing was run, which did still not pass within specifications.The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
Repeatability testing was performed with acceptable results.The issue did not re-occur.A general reagent issue could be excluded.The investigation determined the service actions performed resolved the issue.The issue is consistent with a contaminated sample probe due to improper pre-analytical handling or maintenance by the operator.
|
| |
|