ZIMMER BIOMET, INC. HEADLESS TROCAR DRILL PIN 3.2 MM DIAMETER 75 MM LENGTH; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 07/26/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Initial reporter also sent report to fda: 3800040000-2022-8011.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the pin was placed by the surgeon and when he went to remove it, it fell out.The team was unable to locate the pin on the sterile field at that time and thought it may have fallen by the leg positioning plate.The surgeon was confident that it was not in the patient.During closure, the tech asked the team if they were able to locate the pin and told them they would need to find it before the end of the case.The surgeon was still confident that it was not in the patient.The room was searched and found the pin on the floor.When they inspected it, they found that half of it was broken off.They escalated her concern to the pa and surgeon.An x-ray was taken and the broken piece was visualized in the tibia.It broke off far enough into the bone that it could not be seen without x-ray and would cause more patient harm if it was removed.The surgeon left the broken piece of the pin in place and patient was transported to pacu.While it would have been helpful to find the missing piece of the pin that broke off in the bone sooner, it would not have changed the treatment for the patient.Attempts have been made and no further information has been provided.
 
Event or Problem Description
It was further reported that the patient has no required any further medical intervention.Attempts have been made and all available information has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual examination of the provided pictures identified the unit as broken.The complaint was confirmed.The device history record was reviewed and no discrepancies relevant to the reported event were found.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEADLESS TROCAR DRILL PIN 3.2 MM DIAMETER 75 MM LENGTH
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15904447
Report Number0001822565-2022-03283
Device Sequence Number12043763
Product Code LXH
UDI-Device Identifier00889024201620
UDI-Public(01)00889024201620(17)320613(10)65336445
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00590102000
Device Lot Number65336445
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/07/2022
Supplement Date Received by Manufacturer12/12/2022
03/07/2023
Initial Report FDA Received Date12/02/2022
Supplement Report FDA Received Date12/14/2022
03/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Hospitalization;
Patient Age71 YR
Patient SexFemale
Patient Weight97 KG
-
-