• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH DADE INNOVIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH DADE INNOVIN Back to Search Results
Model Number DADE INNOVIN
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A united states (us) customer contacted a siemens customer care center (ccc).Quality controls (qc) recovered in range and no instrument errors occurred at the time of the event.There were no issues with other samples.A siemens customer service engineer (cse) was dispatched to the customer site and performed a precision run, which recovered acceptably.The patient's clinical condition or pre-analytical variables specific to the first sample collection cannot be ruled out as potential causes of the event.The cause of the event is unknown.The reagent is performing according to specifications.
 
Event or Problem Description
Three falsely elevated prothrombin time (pt) results and three falsely elevated prothrombin time international normalized ratio (inr) results were obtained on a patient sample on a sysmex ca-620 system using dade innovin reagent.A high inr result was reported to the physician(s) and was questioned.The patient was redrawn and was rerun for pt and inr, recovering lower.The lower results were in line with the patient's home meter.The lower inr result was reported, as the correct result, to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the falsely elevated pt and inr results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DADE INNOVIN
Common Device Name
DADE INNOVIN
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
emil-von-behring-str. 76
marburg, D-350 41
GM  D-35041
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
emil-von-behring-str. 76
marburg, D-350 41
GM   D-35041
Manufacturer Contact
christopher aebig
511 benedict ave.
tarrytown, NY 10591
9144153450
MDR Report Key15904526
Report Number9610806-2022-00056
Device Sequence Number6097044
Product Code GJS
UDI-Device Identifier00630414023489
UDI-Public00630414023489
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K974343
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/15/2023
Device Model NumberDADE INNOVIN
Device Catalogue Number10873567
Device Lot Number549787
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/07/2022
Initial Report FDA Received Date12/02/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age73 YR
Patient SexFemale
-
-