• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ANGIOJET SPIROFLEX; CATHETER, EMBOLECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ANGIOJET SPIROFLEX; CATHETER, EMBOLECTOMY Back to Search Results
Model Number 10706
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/09/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that catheter break occurred.An angiojet spiroflex was used for thrombectomy procedure.Upon preparation, the guide catheter was inserted and appeared to be cracked.The procedure was completed with another of the same device.There were no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of the spiroflex thrombectomy.The pump, effluent/supply line, shaft, tip, piston and spike line were visually inspected.Visual examination that the shaft is kinked and the hypotube is separated 40.7cm from the tip.Microscopic examination revealed no additional damages.
 
Event or Problem Description
It was reported that catheter break occurred.An angiojet spiroflex was used for thrombectomy procedure.Upon preparation, the guide catheter was inserted and appeared to be cracked.The procedure was completed with another of the same device.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIOJET SPIROFLEX
Common Device Name
CATHETER, EMBOLECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15905072
Report Number2124215-2022-49356
Device Sequence Number5762841
Product Code MCX
UDI-Device Identifier08714729889380
UDI-Public08714729889380
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K072769
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/27/2022
Device Model Number10706
Device Catalogue Number10706
Device Lot Number0027027301
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/22/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/09/2022
Supplement Date Received by Manufacturer12/08/2022
Initial Report FDA Received Date12/02/2022
Supplement Report FDA Received Date12/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-