| Model Number |
10706 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/09/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that catheter break occurred.An angiojet spiroflex was used for thrombectomy procedure.Upon preparation, the guide catheter was inserted and appeared to be cracked.The procedure was completed with another of the same device.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of the spiroflex thrombectomy.The pump, effluent/supply line, shaft, tip, piston and spike line were visually inspected.Visual examination that the shaft is kinked and the hypotube is separated 40.7cm from the tip.Microscopic examination revealed no additional damages.
|
| |
|
Event or Problem Description
|
|
It was reported that catheter break occurred.An angiojet spiroflex was used for thrombectomy procedure.Upon preparation, the guide catheter was inserted and appeared to be cracked.The procedure was completed with another of the same device.There were no patient complications reported.
|
| |
|