Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a shoulder surgery the device broke off.There was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.Update (b)(6) 2022: the device broke in the patient and everything was removed.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
The complaint is confirmed.One unpackaged ar-4068-25tr serial/batch number (b)(6) was received for investigation.Visual evaluation found that the suturelasso¿ sd, 25° tight curve right tip was broken off.No fragments were returned for analysis.This event occurred before receiving the device for investigation.Additionally, multiple scratches and burrs were observed on the device's shaft.The root cause of the reported failure mode is undetermined.However, this event is most likely caused by prying/levering the device against hard bone or other instrumentation during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUTURELASSO, 25 DEGREE TIGHT CURVE RIGHT
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key15905285
Report Number1220246-2022-05840
Device Sequence Number20357579
Product Code LXH
UDI-Device Identifier00888867031449
UDI-Public00888867031449
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 01/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2026
Device Model NumberSUTURELASSO, 25 DEGREE TIGHT CURVE RIGHT
Device Catalogue NumberAR-4068-25TR
Device Lot Number1032109841
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/14/2022
Supplement Date Received by Manufacturer11/14/2022
Initial Report FDA Received Date12/02/2022
Supplement Report FDA Received Date01/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1