BECTON, DICKINSON & CO. (SPARKS) BD BBL¿ TRYPTICASE¿ SOY BROTH; CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL Back to Search Results
Model Number 297354
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported that bd bbl¿ trypticase¿ soy broth contamination occurred.The following information was provided by the initial reporter: customer states product exhibits low fill volume and contamination.Hazard, injury or erroneous results? no hazard, injury or erroneous results details.
 
Additional Manufacturer Narrative
H.6 investigation summary: material 297354 is manufactured by rehydrating the media components with usp purified water, and thoroughly mixing until a homogeneous solution is obtained.The tubes are filled, capped, torqued, and then labeled by machine per standard operating procedure (sop).The tubes are terminally autoclaved in an air over pressure (aop) autoclave, per manufacturing instructions, using a validated cycle.Post autoclaving, tubes are packaged into final shipping configurations.The batch history record review for batch 2096756 was satisfactory and no quality notifications were generated during manufacturing and inspection.Formulation, filling, and autoclaving processes were within specifications.In process checks were performed at the designated intervals.Those checks confirmed that the caps were tightened to the validated specifications per internal procedure.Qc inspection and testing were satisfactory at time of release.As part of the release criteria for this product, the bhr is reviewed to confirm the following: --the total elapsed time between end of formulation and start of the autoclave cycle was within the specified limits.--all autoclave parameters conformed to the validated cycle parameters for this product.--the minimum f0 for this product was met.The complaint history was reviewed, and no other complaints have been taken for contamination, fill volume, or broken tubes on this batch.Retention samples from batch 2096756 (10 tubes) were available for inspection.No defects were observed in 10/10 retention samples from visual inspection.For further investigation of contamination, two tubes were placed into incubation.One retention tube went into 33-37-degree celsius incubation, and one tube went into 20-25-degree celsius incubation.At the end of a seven-day incubation period no microbial growth was observed.One photo was received to assist with the investigation: ¿ the photo shows four tube from batch 2096756.The media in one tube appears to be turbid and the fill volume appears to be lower than expected in two tubes.The determination of a broken or cracked tube cannot be determined from the photo provided.No returns were received to assist with the investigation.This complaint cannot be confirmed for broken tubes based on the evidence provided by the photo received.This complaint can be confirmed based on the evidence provided by the photo received for contamination, and fill volume.Bd will continue to trend complaints for contamination, fill volume, and broken tubes.Bd ships product in temperature controlled trucks to distribution centers.Distribution centers are provided with shipping and storage guidelines.Mishandling of the product and vibrations during shipping can cause broken tubes.
 
Event or Problem Description
It was reported that bd bbl¿ trypticase¿ soy broth contamination occurred.The following information was provided by the initial reporter: customer states product exhibits low fill volume and contamination.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BBL¿ TRYPTICASE¿ SOY BROTH
Common Device Name
CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15906384
Report Number1119779-2022-01458
Device Sequence Number20277555
Product Code JSG
UDI-Device Identifier30382902973547
UDI-Public30382902973547
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date10/04/2023
Device Model Number297354
Device Catalogue Number297354
Device Lot Number2096756
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/28/2022
Supplement Date Received by Manufacturer01/19/2023
Initial Report FDA Received Date12/02/2022
Supplement Report FDA Received Date01/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/06/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-