• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL (BD WEST) MEDICAL SURGICAL BD INTERLINK¿ VIAL ACCESS CANNULA; HYPODERMIC SINGLE LUMEN NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL (BD WEST) MEDICAL SURGICAL BD INTERLINK¿ VIAL ACCESS CANNULA; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Model Number 303367
Medical Device Problem Codes Leak/Splash (1354); Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the bd interlink¿ vial access cannula dislodged the stopper into the vial causing leakage.The following information was provided by the initial reporter: when accessing a medication vial with the twin-pack replacements instead of puncturing the rubber plug, it punches it into the vial and the medication pours out.
 
Additional Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device expiration date: unknown.Device manufacture date: unknown.
 
Event or Problem Description
It was reported that the bd interlink¿ vial access cannula dislodged the stopper into the vial causing leakage.The following information was provided by the initial reporter: when accessing a medication vial with the twin-pack replacements instead of puncturing the rubber plug, it punches it into the vial and the medication pours out.
 
Additional Manufacturer Narrative
Investigation summary as no physical sample, picture sample, or lot number was provided for evaluation by our quality engineer team, a complete investigation could not be performed.A device history review could not be completed as no batch number was provided.Based on the limited investigation results, a cause for the reported incident could not be determined.
 
Event or Problem Description
It was reported that the bd interlink¿ vial access cannula dislodged the stopper into the vial causing leakage.The following information was provided by the initial reporter: when accessing a medication vial with the twin-pack replacements instead of puncturing the rubber plug, it punches it into the vial and the medication pours out.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval? yes.D10: returned to manufacturer on: 19-dec-2022.H6.Investigation summary: it was reported there were multiple issues when using the product.To aid in the investigation, one sample with no packaging blister was received for evaluation by our quality team.A visual inspection was performed and no defects or imperfections were observed.As the lot provided is 'unknown,' a device history record review could not be completed.Based on the investigation and with the returned sample analysis the symptom reported by the customer could not be confirmed and a probable root cause could not be offered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INTERLINK¿ VIAL ACCESS CANNULA
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15906873
Report Number1911916-2022-00692
Device Sequence Number5391062
Product Code FMI
UDI-Device Identifier30382903033677
UDI-Public(01)30382903033677
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K924439
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number303367
Device Catalogue Number303367
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/07/2022
Supplement Date Received by Manufacturer12/15/2022
01/09/2023
Initial Report FDA Received Date12/02/2022
Supplement Report FDA Received Date01/03/2023
01/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-