Model Number 0.805.9510
Medical Device Problem Code Unintended Electrical Shock (4018)
Health Effect - Clinical Code Electric Shock (2554)
Date of Event 11/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A dealer technician was on site 11/7/2022 at the doctor's office and checked the unit for visual evidence of a possible electrical short (scorch marks), and for any exposed wires on the unit.There were no evidence of an electrical short, nor any exposed wires found.The dealer technician took multiple exposures, whereby no errors or malfunction occurred.On 11/17/2022 a dental imaging technologies corporation (ditc) technician (initial importer) visited the doctor's office to take photos of the device and perform additional testing.The incoming voltage to the unit was constant.In review of the photos, two exposed wires were observed.Tube head ground wire.Scissor arm cable where it meets the horizontal arm.The technician did not see any other cable damage inside the scissor or horizontal arm, or on the generator board.The staff member who reported being shocked was not in the office during this visit, thus the actual location of where the reported spark occurred could not be confirmed.The unit was being used at time of this visit and the ditc technician advised the manager to discontinue use of the unit, and applied appropriate signage on the unit.The unit will be returned to the manufacturer for further investigation, whereby a follow-up report will be submitted when the investigation is completed.
 
Event or Problem Description
It was reported by a dental office that a staff member received a shock from the unit while preparing for clinical use.Staff reported seeing a white spark/ flash from the back of the horizontal arm where it meets the wall.No injuries or medical attention was reported.
 
Additional Manufacturer Narrative
The unit investigation was completed on 1/13/2023.A broken tubehead protective earthing cable was discovered, whereby two (2) strands were broken.There is no evidence of the cause of the broken strands.The electrical safety test was run for the device in production.All test cases were passed except the tubehead arm earth resistance test due to broken pe wire.Even with broken wire, the tubehead arm earth resistance was still only 101mohm, while the highest allowed value in production is 90mohm.The safety standard 60601-1 allows 100mohm resistance.All other cables, cable joints and connectors were in good condition, with no signs of corrosion, sparking or burn anywhere on the device.The reported problem i.E.Electric shock was not repeatable and no reason was found that could have caused the electric shock.Based on the investigations the product has stayed safe despite the broken pe wire.This concludes the investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOCUS
Common Device Name
EXTRAORAL SOURCE X-RAY SYSTEM
Manufacturer (Section D)
PALODEX GROUP OY
nahkelantie 160
tuusula, uusimaa 04300
FI  04300
Manufacturer (Section G)
PALODEX GROUP OY
nahkelantie 160
tuusula, uusimaa 04300
FI   04300
Manufacturer Contact
rainer harjunpaa
nahkelantie 160
tuusula, uusimaa 04300
FI   04300
MDR Report Key15907048
Report Number3005383085-2022-00003
Device Sequence Number5391066
Product Code EHD
UDI-Device Identifier06430035872695
UDI-Public(01)06430035872695
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K001729
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 12/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0.805.9510
Device Catalogue Number900986
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/07/2022
Supplement Date Received by Manufacturer11/07/2022
Initial Report FDA Received Date12/02/2022
Supplement Report FDA Received Date02/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/13/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-