| Model Number |
AESR60R |
| Medical Device Problem Code |
Fail-Safe Did Not Operate (4046)
|
| Health Effect - Clinical Code |
Gastrointestinal Hemorrhage (4476)
|
| Date of Event |
10/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
An aeon endostapler 3.25mm/60mm orange reload was used with an aeon endostapler handle - long in a sleeve gastrectomy procedure.The procedure took place on (b)(6) 2022 and lexington medical was made aware of the event on november 2, 2022.The stapler cut but did not staple.No patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
After investigation, it was determined that the level one root cause for no staples present after retracting was that the reload had been re-fired.The level two root cause was determined to be a combination of user error and product malfunction.Reloads are single-use devices that are not intended to be re-fired, as per the warning in the instructions for use 900299 section 1.4.9.However, a single-use lockout is present on all reloads to prevent re-firing of used reloads.
|
| |
|