| Model Number |
9-ASD-012 |
| Medical Device Problem Code |
Difficult to Fold, Unfold or Collapse (1254)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2022, a 12mm amplatzer septal occluder was selected for implantation using an unknown abbott delivery system.During the procedure, during deployment, the device deformed into a cobra shape.The device was removed from patient anatomy and replaced with a another 12mm amplatzer duct occluder.The device was not released inside patient anatomy, there was no angulation or kink noted in the delivery system, and there was no interaction with cardiac structures during deployment.The patient is reported to be stable.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of device deformity was reported.The device was returned to abbott for investigation and the device met functional specifications when analyzed under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.The cause of the reported event could not be conclusively determined.
|
| |
|