Model Number 1320-3042
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 10/03/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.Should additional information become available, it will be provided in a supplemental report.Device disposition unknown.
 
Event or Problem Description
As reported: "during the procedure, a piece of drill bit got stuck in the patient's femoral shaft, whose bone quality was poor.The surgeon preferred not to insist on this procedure because of the risk of damaging the bone, given the patient's life expectancy.Because the piece of drill bit remained in the patient, the event is considered a serious event." the patient had metastases and a very hard bone.In no case it is a problem of the drill.The fact that a piece of drill remains in the patient, is not serious.
 
Event or Problem Description
As reported: "during the procedure, a piece of drill bit got stuck in the patient's femoral shaft, whose bone quality was poor.The surgeon preferred not to insist on this procedure because of the risk of damaging the bone, given the patient's life expectancy.Because the piece of drill bit remained in the patient, the event is considered a serious event." the patient had metastases and a very hard bone.In no case it is a problem of the drill.The fact that a piece of drill remains in the patient, is not serious.
 
Additional Manufacturer Narrative
The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on investigation and the provided information, the root cause was attributed to a patient related issue.The available information indicates failure was caused by a mechanical overload caused by very hard bone related to metastases.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL, AO SMALL GAMMA3® Ø4.2X300MM
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15911569
Report Number0009610622-2022-00568
Device Sequence Number19462788
Product Code LXH
UDI-Device Identifier04546540377883
UDI-Public04546540377883
Combination Product (Y/N)N
Initial Reporter CountrySZ
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1320-3042
Device Catalogue Number13203042
Device Lot NumberKU105559
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/09/2022
Supplement Date Received by Manufacturer01/10/2023
Initial Report FDA Received Date12/02/2022
Supplement Report FDA Received Date02/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/20/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-