| Model Number |
TMICL 12.1 |
| Medical Device Problem Codes |
Off-Label Use (1494); Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/26/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that a 12.1mm tmicl 12.1 implantable collamer lens of -9.0/3.0/102 (sphere/cylinder/axis) was implanted into the patients left eye (os) on (b)(6) 2022.Low vault was observed.On (b)(6) 2022 the lens was exchanged for a longer length lens.Cause reported as other."lens too short." post operative information is not available.If additional information is received a supplemental medwatch report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5-the reporter indicated that a 12.1mm tmicl 12.1 implantable collamer lens of -9.0/3.0/102 (sphere/cylinder/axis) was implanted into the patients left eye (os) on 23-feb-2022.Low vault was observed.On 30-nov-2022 the lens was exchanged for a longer length lens.The exchange resolved the problem.Cause reported as other."lens too short." corrected data: d4: correct serial number: t946635.Expiration date: 31-jan-2024.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Claim#: (b)(4).
|
| |
|