| Model Number |
23375 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 25-nov-2022.It was reported that balloon damage occurred.The patient presented with may-thurner syndrome and underwent venous angioplasty with intervention.The stenosed target lesion was located in the iliac vein.A 16-2/5.8/75 xxl esophageal balloon catheter was advanced for dilatation.However, during inflation, the balloon kept losing pressure despite disconnecting and reconnecting the indeflator.The procedure was completed with another of same device.No patient complications were reported.However, returned device analysis revealed a balloon pinhole.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: the device was received for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.Blood present in balloon which is indicative of a leak.The balloon was attached to an encore inflation device, subjected to positive pressure and liquid was observed to be leaking from a balloon pinhole located 30mm distal of the proximal balloon sleeve.An examination of the balloon material and the tip region identified no issues which could potentially have contributed to this complaint.The rated burst pressure for this device is 5 atmospheres as per xxl specification.A visual and tactile examination identified no kinks or damage to the shaft which may have potentially contributed to the complaint incident.A visual examination identified no issues with the tip of the device that could have contributed to the complaint incident.
|
| |
|
Search Alerts/Recalls
|