• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT; BZE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT; BZE Back to Search Results
Model Number RT380
Device Problems Break (1069); Leak/Splash (1354); Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/14/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The complaint rt380 adult dual heated evaqua2 breathing circuit is currently en route to fisher & paykel healthcare (f&p) for evaluation.We will provide a follow-up report upon completion of our investigation.
 
Event Description
A distributor reported on behalf of a healthcare facility in japan that the tubing of a rt380 adult dual heated evaqua2 breathing circuit was found cracked.There was no reported patient involvement.
 
Event Description
A distributor reported on behalf of a healthcare facility in japan that the tubing of a rt380 adult dual heated evaqua2 breathing circuit was found cracked.There was no reported patient involvement.
 
Manufacturer Narrative
(b)(4).Method: the complaint rt380 adult dual heated evaqua2 breathing circuit was returned to fisher & paykel healthcare (f&p) in new zealand, where it was visually inspected and analysed.Results: visual inspection of the returned circuit revealed that the expiratory tube of a rt380 adult dual heated evaqua2 breathing circuit was degraded.Further testing revealed that the tubing had been exposed to chemicals that induced hydrolysis and degradation.Conclusion: we are unable to determine the cause of the reported event.However, based on our investigation, it was most likely due to chemical exposure to the expiratory limb.All rt380 adult dual-heated evaqua2 breathing circuits are visually inspected, and pressure and flow tested during production, and those that fail are rejected.The subject breathing circuit would have met the required specifications at the time of production.The user instructions that accompany the rt380 breathing circuit state: "do not soak, wash, sterilize or reuse this product.Avoid contact with chemicals, cleaning agents or hand sanitizers." "do not use medications containing tyloxapol (such as tacholiquin) as this may damage the tubing and lead to a loss of ventilation pressure." "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." "ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT
Type of Device
BZE
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
17400 laguna canyon road
suite 300
irvine, CA 92618
9494534000
MDR Report Key15912662
MDR Text Key304803491
Report Number9611451-2022-01132
Device Sequence Number1
Product Code BZE
UDI-Device Identifier09420012429728
UDI-Public(01)09420012429728(10)2102069994(11)220307
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K103767
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/04/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRT380
Device Catalogue NumberRT380
Device Lot Number2102069994
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received03/31/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/07/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ASTRAL VENTILATOR
-
-