| Model Number |
405108 |
| Medical Device Problem Codes |
Difficult to Insert (1316); Activation, Positioning or Separation Problem (2906)
|
| Health Effect - Clinical Codes |
Cardiac Perforation (2513); Vascular Dissection (3160)
|
| Date of Event |
11/09/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
During the procedure, the introducer needle was not moving which led to a perforation in the subclavian vein.
|
| |
|
Additional Manufacturer Narrative
|
|
Further information was received confirming the perforation was diagnosed using fluoroscopy and the procedure was completed.The patient is stable.
|
| |
|
Event or Problem Description
|
|
During the rv implant procedure, the introducer needle was not moving in the sheath which led to a perforation in the subclavian vein.The perforation was diagnosed using fluoroscopy when the catheter was being advanced.The procedure was successfully completed and the patient is stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected information: h6.
|
| |
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|