• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE (127 DEG) SIZE 4.5 ACCOLADE (127 DEG) SIZE 4.5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE (127 DEG) SIZE 4.5 ACCOLADE (127 DEG) SIZE 4.5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 6021-4535
Medical Device Problem Codes Degraded (1153); Unstable (1667)
Health Effect - Clinical Codes Pain (1994); Joint Laxity (4526)
Date of Event 11/15/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.Device not returned to the manufacturer.
 
Event or Problem Description
Patient was implanted with an lfit cocr v40 femoral head on her left hip on (b)(6) 2010.It is further alleged that she noted increasing pain and feeling of instability in her hip.X-rays obtained in october allegedly showed trunnion damage and the head on an angle.During revision surgery on (b)(6) 2019, "the ball was grossly loose on the trunnion and"."the tip of the trunnion had been whittled down to a sharp point." "there was gross damage from the sharp trunnion tip riding into the polyethylene.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE (127 DEG) SIZE 4.5 ACCOLADE (127 DEG) SIZE 4.5
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
arokiya raj
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15913162
Report Number0002249697-2022-01769
Device Sequence Number6562919
Product Code LPH
UDI-Device Identifier04546540510761
UDI-Public04546540510761
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2019
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2015
Device Model Number6021-4535
Device Catalogue Number6021-4535
Device Lot Number33149202
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/10/2022
Initial Report FDA Received Date12/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/13/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age84 YR
Patient SexFemale
-
-