• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LUMENIS BE LTD. LIGHSHEER DUET; SOLID STATE AESTHECIC LASER DELIVERY DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LUMENIS BE LTD. LIGHSHEER DUET; SOLID STATE AESTHECIC LASER DELIVERY DEVICE Back to Search Results
Model Number LIGHTSHEER DUET HS/ET
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Burn(s) (1757)
Date of Event 11/08/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Lumenis investigated the reported event by contacting the user facility directly.Attempts by emails and phone calls have been made to obtain, patient treatment settings, patient information, patient photo.The device gasb00000:(b)(4) was installed in the facility on (b)(6) 2013 and the last pm was done on (b)(6) 2021.Service engineer visited the facility and tested the device.He checked all the power on both handpieces and verified that output is within specification.Performed power detector calibration per 7.3 of service manual.And checked all power levels per 7.5 of service manual.No failure found.Device malfunction wasn't the suspected cause of the adverse event.According to the information received there is no malfunction found on a system.Since the customer neither reply us nor sent us filled out incident form, lumenis is considering this case as a safety one, like if it was supposed to cause serious injury with no permanent impairment and reporting given event in an 'abundance of caution'.Should additional significant information become available, the complaint record will be updated accordingly and a follow-up medwatch report will be submitted.
 
Event or Problem Description
Lumenis received an adverse event report on a patient who sustained burn following treatment by ls duet device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGHSHEER DUET
Common Device Name
SOLID STATE AESTHECIC LASER DELIVERY DEVICE
Manufacturer (Section D)
LUMENIS BE LTD.
6 hakidma st.
po box 240
yokneam, isr 20692 04
IS  2069204
Manufacturer (Section G)
LUMENIS BE LTD.
6 hakidma st.
po box 240
yokneam, isr 20692 04
IS   2069204
Manufacturer Contact
roman vernik
6 hakidma st.
po box 240
yokneam, isr 20692-04
IS   2069204
MDR Report Key15914068
Report Number3021349626-2022-00005
Device Sequence Number17239336
Product Code GEX
UDI-Device Identifier07290109140384
UDI-Public07290109140384
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K170179
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Administrator/Supervisor
Remedial Action Inspection
Type of Report Initial
Report Date (Section B) 12/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberLIGHTSHEER DUET HS/ET
Device Catalogue NumberGASB00000
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/08/2022
Initial Report FDA Received Date12/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/11/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-