ARTHREX, INC. CLAV FRAC PLT,CENTRAL THIRD,LFT,SS, STRL; PLATE, FIXATION, BONE Back to Search Results
Model Number CLAV FRAC PLT,CENTRAL THIRD,LFT,SS, STRL
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a few weeks after the surgery the clavicula plate broke in two parts in the patient.Update 17-nov-2022: the initial surgery took place on (b)(6) 2022, the breakage of the plate was noticed on (b)(6) 2022 and treated on (b)(6) 2022.The revision surgery was performed with two clavicle plates and cancellous bone.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
Visual inspection found that the device was broken in half.Possible cause cannot be determined with the given information but likely causes include excessive force on the device during procedure, trauma to the patient, and or device related failure.Complaint was confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAV FRAC PLT,CENTRAL THIRD,LFT,SS, STRL
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key15914093
Report Number1220246-2022-05857
Device Sequence Number4459662
Product Code HRS
UDI-Device Identifier00888867195929
UDI-Public00888867195929
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K143139
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 01/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberCLAV FRAC PLT,CENTRAL THIRD,LFT,SS, STRL
Device Catalogue NumberAR-2653CLS
Device Lot Number14921868
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/17/2022
Supplement Date Received by Manufacturer11/17/2022
Initial Report FDA Received Date12/05/2022
Supplement Report FDA Received Date01/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-