Model Number BONE CUTTER,5.5MM X 13CM
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 11/17/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported on (b)(6) 2022 by a sales representative via sems that an ar-8550bc bone saw proximal end broke off shaver "sounded like it melted off".Case involvement, device broke during use.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
The complaint is confirmed.The reported device was not returned for investigation.No pictures were provided.However, per the event description is concluded that the most likely cause(s) of this type of event include the device had been bent by applying excessive force when prying and/or leveraging it (side-loading).The torque produced might have caused the plastic hub, the interface with the motor rotating parts, to deform and break.This characterizes misuse.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BONE CUTTER,5.5MM X 13CM
Common Device Name
BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key15914340
Report Number1220246-2022-05867
Device Sequence Number17258050
Product Code GFA
UDI-Device Identifier10888867043937
UDI-Public00888867043930
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 06/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBONE CUTTER,5.5MM X 13CM
Device Catalogue NumberAR-8550BC
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/17/2022
Supplement Date Received by Manufacturer11/17/2022
Initial Report FDA Received Date12/05/2022
Supplement Report FDA Received Date06/07/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-