• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ PANOPTIX UV ABSORBING TRIFOCAL IOL; LENS, MULTIFOCAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ PANOPTIX UV ABSORBING TRIFOCAL IOL; LENS, MULTIFOCAL Back to Search Results
Model Number TFAT00
Medical Device Problem Codes Application Program Problem: Power Calculation Error (1473); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Visual Disturbances (2140)
Date of Event 06/23/2021
Type of Reportable Event Serious Injury
Event or Problem Description
A facility representative reported that following an intraocular lens (iol) implant procedure, the patient experienced poor quality vision in distance and near.The iol was exchanged for monofocal lens model 17 months following the initial implant procedure.The clinical reason for explant mentioned as unknown reason and attempted all treatments.Additional information has been requested.
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been one other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
Additional information received and stated that the patient was not hospitalized for the event and there was no patient harm.Also stated that, aggressive treatment for dry eye, attempts at refraction was done.
 
Additional Manufacturer Narrative
Additional information provided in b.3., b.5., b.6., b.7., d.10., e.4., h.3., h.6., and h.10.The account indicated the use of a qualified cartridge and handpiece.The viscoelastic used is not qualified for use with any qualified lens/cartridge/handpiece combination.The product investigation could not identify a root cause for the reported complaint.Each lens is subjected to a 100% assessment of the power and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Information was provided that the multifocal company 18.0 diopter (add power +2.2/+3.2) lens was replaced with a monofocal company model 18.0 diopter lens.The product was not available for evaluation.Due to the exchange of a multifocal lens to a monofocal lens, this may suggest that the replacement lens model was an improved choice for the patient's vision needs.No additional information has been provided at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ PANOPTIX UV ABSORBING TRIFOCAL IOL
Common Device Name
LENS, MULTIFOCAL
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15914451
Report Number1119421-2022-02561
Device Sequence Number17823332
Product Code MFK
UDI-Device Identifier00380652429672
UDI-Public00380652429672
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P040020
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2024
Device Model NumberTFAT00
Device Catalogue NumberTFAT00.180
Device Lot Number12740987
Was Device Available for Evaluation? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/09/2022
Supplement Date Received by Manufacturer12/08/2022
Initial Report FDA Received Date12/05/2022
Supplement Report FDA Received Date01/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/13/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH III CARTRIDGE D; MONARCH III INJECTOR; NUVISC; ORA SYSTEM
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexUnknown
-
-