Model Number 1860
Medical Device Problem Code Component Missing (2306)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluated by mfr: returned product consisted of an fg threader mr 1.2mm x 12mm catheter.Device was received inside boxed packaging and pouch; however, pouch seal was broken, and carton was taped closed with tape.Visual observation of the device identified that shaft separation at 129.5 cm from the strain relief.The distal inflation lumen from the separation point was missing from the returned device.Numerous kinks were identified on the hypotube.Product analysis confirms the reported event as balloon, distal tip, and inflation lumen from the separation point was identified missing.
 
Event or Problem Description
Reportable based on device analysis completed on (b)(6) 2022.It was reported that a device component was missing.Upon unpacking of a 1.2mm x 12mm threader balloon catheter, it was noticed that the balloon was missing.There were no patient complications reported.However, returned device analysis revealed a shaft detached/separated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THREADER
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15914767
Report Number2124215-2022-50495
Device Sequence Number15152942
Product Code LOX
UDI-Device Identifier08714729848127
UDI-Public08714729848127
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K134031
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/07/2023
Device Model Number1860
Device Catalogue Number1860
Device Lot Number0027619886
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/14/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/23/2022
Initial Report FDA Received Date12/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age72 YR
Patient SexMale
Patient Weight60 KG
-
-