MAUDE Adverse Event Report: MEDTRONIC IRELAND ONYX FRONTIER; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND ONYX FRONTIER; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number ONYXNG22512UX
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/23/2022
Type of Reportable Event Serious Injury
Event or Problem Description
An attempt was made to use one onyx frontier drug eluting stent to treat a severly calcified lesion with 80% stenosis in the distal obtuse marginal (om).The device was inspected with no issues noted.Negative prep was not performed.The lesion was pre-dilated, the device did pass through a previously deployed stent.Resistance was encountered when advancing the device however excessive force was not used during delivery.It was reported that stent dislodgement occurred during delivery to/at lesion.The dislodged stent was not removed.It was stated that wire placement was lost.The family were spoken to with respect to surgery options.The patient is reported to be alive with no injury.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Additional information: a non mdt wire was advanced to the distal om branch for support.Following pre-dilation using a 2.5 x 15 mm balloon the lesion was stented with the 2.5 x 22 non medtronic stent.Because of signicicant residual stenosis just distal to this stent the decision was made to stent this area.Upon advancing a 2.25x22mm onyx frontier drug eluting stent, however upon advancing the stent, the stent detached, from the stent delivery system within the previously placed stent likely after hanging up on calcium.While attempting to advance a balloon catheter over the wire through the dislodged stent, the entire guide catheter system was dislodged from the left main coronary artery.It was stated that wire placement was lost.With significant difficulty, a wire was re-advanced into the distal om branch after re-engagement of the left main coronary artery with a guide catheter.Despite the use of multiple techniques and equipment, a balloon catheter could not pass by the dislodged stent.The dislodged stent did not adhere to the vessel wall.The family did not want to proceed with open heart surgery/bypass grafting, despite the risk of acute stent thrombosis.It was decided to pursue with conservative treatment the decision was made to attempt a right coronary artery (rca) pci.A 6f non medtronic guide catheter without side holes was advanced over a wire and used to selectively cannulate the right coronary artery (rca) a non mdt wire was advanced into the distal rca system for support.Sequential balloon dilatation was performed using a 2.0 x 8 mm and 2.5 x 12 mm balloon with minimal luminal gain due to severe concentric calcification.Intravascular ultrasound was to be performed but the catheter could not be advanced past the severe stenosis of the proximal segment.Additional balloon dilatation was performed without significant luminal gain, an intraluminal coronary lithotripsy was attempted using the shockwave system however the non mdt balloon could not pass the proximal lesion.Multiple attempts were then performed without significant improvement.Due to the length of the procedure and the amount of contrast used, the decision was made to stop the intervention.B1: updated from product problem to adverse event and product problem.H1: updated from malfunction to serious injury medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ONYX FRONTIER
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key15914769
Report Number9612164-2022-04574
Device Sequence Number6798088
Product Code NIQ
UDI-Device Identifier00763000511197
UDI-Public00763000511197
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
P160043
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/11/2024
Device Model NumberONYXNG22512UX
Device Catalogue NumberONYXNG22512UX
Device Lot Number0011067719
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/30/2022
Supplement Date Received by Manufacturer12/19/2022
Initial Report FDA Received Date12/05/2022
Supplement Report FDA Received Date01/12/2023
Date Device Manufactured02/11/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age79 YR
Patient SexMale
Patient Weight88 KG
-
-