| Model Number |
IPN913391 |
| Medical Device Problem Code |
Activation Problem (4042)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
11/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).The device investigation is pending, and the results will be submitted in a follow-up report.Teleflex will continue to monitor and trend related events.
|
| |
|
Event or Problem Description
|
|
It was reported that the balloon was defective during use.There was no reported injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).The defective sample was returned but the reported defect cannot be confirmed based on visual inspection.A dimensional inspection is not relevant to complained defect.The blue balloon was inflated by air and no leakage or cross-leak between balloons was observed.Yellow balloon was inflated by air.Within 30s the balloon fully deflated.The reported defect can be confirmed based on functional inspection.A leak was observed on the yellow balloon.The defective complained catheter cat-02770-002 was send to the supplier for detail investigation.There was identified that the cylindrical part of the flexible part of the valve is not perfectly adhered to the valve.There is a tunnel on the neck through which air escapes from the valve.It looks like a valve manufacturing defect.From this reason, teleflex hradec decided to initiate an nc.Subsequently, the valve supplier, teleflex malaysia was contacted.All complaints are trended across product families on a monthly basis according to the documented standards.As the lot number of the defective product was reported, the device history review of the complained lot could be completed.Final inspection records of used catheters cat-02770-003 (lots 72c21k0438, 72c21k0203) were reviewed.Lots were checked by final qc inspection in hradec and were released without any defect observed.The review of device history revealed no production issues from the perspective of the reported defect and the time of manufacture of the complained lot.The root cause of this issue is supplier related.The issue comes from hradec facility manufacturing process.An nc was opened on supplier side to identify a root cause of complained defect.In conclusion, the customer complaint regarding the ez-blocker product was reported.As the lot number of the defective product was reported, the device history review was completed.The review of the device review revealed no production issues at the time of manufacturing of the complained lot.The complained defect was evaluated with cooperation with supplier and confirmed.There was observed a tunnel on the neck through which air escapes from the yellow valve.No corrective or preventive actions in production are deemed necessary to introduce because the issue was not caused in our manufacturing process.Probable corrective action will be implemented by supplier under the nc.
|
| |
|
Event or Problem Description
|
|
It was reported that the balloon was defecive during use.There was no reported injury.
|
| |
|
Search Alerts/Recalls
|