| Catalog Number |
230H4001 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that the surgeon was final tightening a set screw when the tip of the set screw driver broke off inside of the anchor.The surgeon was able to retrieve the broken driver tip without harm to the patient.
|
| |
|
Event or Problem Description
|
|
It was reported that the surgeon was final tightening a set screw when the tip of the set screw driver broke off inside of the anchor.The surgeon was able to retrieve the broken driver tip without harm to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrections in h3.Additional information in h6: component, investigation type, findings, and conclusions.Device evaluation: visual inspection revealed that the tip of the driver was fractured off.Potential cause: root cause was unable to be determined.This event could possibly be attributed to wear through use over time or multiple sterilization cycles.It could also be attributed to off-axis forces applied to cause the tip to fracture.Dhr review and related actions per dhr review, the returned part was likely conforming when it left zimvie control.Device use : this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|