Model Number C701
Medical Device Problem Codes Incorrect Measurement (1383); Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/16/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The field service engineer found the rinse tubing was damaged and caused a leak into the cuvettes.He replaced the damaged tubing, performed baseline checks, adjusted the sample probe, and adjusted the fluidic dispense level to specification.He performed precision testing with results within specification.The customer ran calibration and qc with results meeting specifications.The investigation is ongoing.
 
Event or Problem Description
There was an allegation of questionable gluc hk gen.3 (glucose) results from the cobas 8000 cobas c 701 module.One example was provided of an initial result of 16.8 mg/dl and a repeat result of 129.2 mg/dl.The questionable results were not reported outside of the laboratory.The repeat results were deemed correct.The reagent lot number and expiration date were requested but were not provided.
 
Additional Manufacturer Narrative
The investigation determined the service actions resolved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS 8000 C 701 MODULE
Common Device Name
CLINICAL CHEMISTRY ANALYZER
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
HITACHI HIGH TECH CORP.
882 ichige hitachinaka
na
ibaraki 312-8 504
JA   312-8504
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15914976
Report Number1823260-2022-03860
Device Sequence Number6544301
Product Code JJE
UDI-Device Identifier04015630925025
UDI-Public04015630925025
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K100853
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberC701
Device Catalogue Number05641489001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/16/2022
Supplement Date Received by Manufacturer12/23/2022
Initial Report FDA Received Date12/05/2022
Supplement Report FDA Received Date12/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-