| Model Number |
G52226 |
| Medical Device Problem Code |
Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is still pending.A follow up mdr will be submitted to include the investigation conclusions.
|
| |
|
Event or Problem Description
|
|
Patient (b)(4) stent #2 - user error: patient received chemotherapy after stent placement.Stent #2: (b)(4) (b)(6) procedure: were any additional procedure (s) performed during the study procedure? no.Were any adverse events documented during the stent procedure? yes.Were any device deficiencies documented during the placement procedure? yes.Was tumor-reduction therapy documented during follow up? chemotherapy.User error: after stent placement, additional methods of treatment such as chemotherapy and irradiation may increase the risk of stent migration due to tumor shrinkage, stent erosion, and/or mucosal bleeding.Patient outcome: at the conclusion of follow-up what was the status of the study stent: stent removed endoscopically.Days post stent placement: 199.Reason for removal: other (unrelated to ae) new stenosis because of hyperplastic tumor tissue.Primary reason for study exit : death.Cause of death: tumor progress.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: user/use related complaints are considered foreseen misuse.It is unknown how the device will perform outside of instructions for use and/or labelling requirements.The user has not complied with the requirements of the ifu and/label.Trending will monitor if any future investigation is required.The device evaluation could not be completed as the device or photographic evidence of the device was not returned for evaluation.Manufacturing records: prior to distribution all devices are subjected to a visual inspection and functional inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.A review of the manufacturing records for evo-fc-20-25-10-e device of lot number c1509514 did not reveal any discrepancies that could have contributed to this issue.There is no evidence to suggest that this issue affects the entire lot #c1509514; upon review of complaints this failure mode has not occurred previously with this lot #c1509514.The instructions for use, which accompanies this device, states ¿after stent placement, alternative methods of treatment such as chemotherapy and radiation should not be administered at this may increase risk of stent migration due to tumour shrinkage, stent erosion, and/or mucosal bleeding.¿¿ there is evidence to suggest that the customer did not follow the instructions for use.Root cause analysis a definitive root cause of user error was established.The user has not complied with the requirements of the ifu and/label.According to the information available the patient was treated with chemotherapy after stent placement.Summary of investigation complaint is confirmed based on customer and/or rep testimony.According to the initial reporter, the patient did experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
|
| |
|
Event or Problem Description
|
|
Supplemental report is being submitted due to the completion of the investigation on 26-jan-23.
|
| |
|
Event or Problem Description
|
|
This correction supplemental report is being submitted because section f10 health effect - impact code (annex f) was updated from f19 to f26 on 08-mar-2023.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: user/use related complaints are considered foreseen misuse.It is unknown how the device will perform outside of instructions for use and/or labelling requirements.The user has not complied with the requirements of the ifu and/label.Trending will monitor if any future investigation is required.The device evaluation could not be completed as the device or photographic evidence of the device was not returned for evaluation.Manufacturing records: prior to distribution all devices are subjected to a visual inspection and functional inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.A review of the manufacturing records for evo-fc-20-25-10-e device of lot number c1440187 did not reveal any discrepancies that could have contributed to this issue.There is no evidence to suggest that this issue affects the entire lot c1440187 ; upon review of complaints this failure mode has not occurred previously with this lot c1440187.The instructions for use, ifu0061 which accompanies this device, states ¿after stent placement, alternative methods of treatment such as chemotherapy and radiation should not be administered at this may increase risk of stent migration due to tumour shrinkage, stent erosion, and/or mucosal bleeding.¿¿ there is evidence to suggest that the customer did not follow the instructions for use.Image review: an image was not returned for evaluation.Root cause analysis: a definitive root cause of user error was established.The user has not complied with the requirements of the ifu and/label.According to the information available the patient was treated with chemotherapy after stent placement.Summary of investigation: complaint is confirmed based on customer and/or rep testimony.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
|
| |
|
Additional Manufacturer Narrative
|
|
Supplemental correction report is being submitted following a review of this complaint file.Correction report was required due to 'correction mdr to be sent as lot # was incorrect in final report'.Device evaluation: user/use related complaints are considered foreseen misuse.It is unknown how the device will perform outside of instructions for use and/or labelling requirements.The user has not complied with the requirements of the ifu and/label.Trending will monitor if any future investigation is required.The device evaluation could not be completed as the device or photographic evidence of the device was not returned for evaluation.Manufacturing records: prior to distribution all devices are subjected to a visual inspection and functional inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.A review of the manufacturing records for evo-fc-20-25-10-e device of lot number c1440187 did not reveal any discrepancies that could have contributed to this issue.There is no evidence to suggest that this issue affects the entire lot c1440187 ; upon review of complaints this failure mode has not occurred previously with this lot c1440187.The instructions for use, ifu0061 which accompanies this device, states ¿after stent placement, alternative methods of treatment such as chemotherapy and radiation should not be administered at this may increase risk of stent migration due to tumour shrinkage, stent erosion, and/or mucosal bleeding.¿¿ there is evidence to suggest that the customer did not follow the instructions for use.Image review: an image was not returned for evaluation.Root cause analysis a definitive root cause of user error was established.The user has not complied with the requirements of the ifu and/label.According to the information available the patient was treated with chemotherapy after stent placement.Summary of investigation complaint is confirmed based on customer and/or rep testimony.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
|
| |
|
Event or Problem Description
|
|
Supplemental correction report is being submitted following a review of this complaint file.Correction report was required due to 'correction mdr to be sent as lot # was incorrect in final report'.
|
| |
|
Search Alerts/Recalls
|