Patient allegedly received an implant on (b)(6) 2013 due to deep vein thrombosis (dvt).Patient is alleging vena cava perforation.Patient alleges "i live with the anxiety of having a filter that could fail further at any time, but that cannot be retrieved without a serious surgery.I am worried because my filter has perforated outside of my vena cava." "for fear of aggravating my current condition and causing potential future injuries related to the cook filter, i try not to engage in any activities that could cause me to exert myself.This includes everyday lifting." per computed tomography report, "caval perforation: yes.10 o'clock 7.00 mm grade 2.".
|
|
The previous medwatch report was submitted by william cook europe under manufacturer report reference# 3002808486-2022-00396.Additional information provided determined that this device was manufactured by cook inc.With the submission of this initial medwatch report, cook inc.Informs that all future submissions regarding this complaint will be handled under the manufacturer report reference # referenced in this initial medwatch report.Investigation is reopened due to additional information provided.The following allegations have been investigated: vena cava (vc) perforation, anxiety, worry, fear, limited physical activity.The reported allegations have been further investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: vascular trauma, vena cava perforation, vena cava penetration.Unknown if the reported anxiety, worry, fear, and limited physical activity are directly related to the filter and unable to identify a corresponding failure mode at this point in time.A total of (b)(4) devices were manufactured in the reported lot.To date, one other complaint has been reported against the lot 4202162.The associated work order was reviewed.No related/relevant notes were documented.The device is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|