ABBOTT VASCULAR NC TREK CORONARY DILATATION CATHETER Back to Search Results
Model Number 1012449-20
Medical Device Problem Codes Material Separation (1562); Improper or Incorrect Procedure or Method (2017); Deformation Due to Compressive Stress (2889); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/07/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.(b)(4).
 
Event or Problem Description
It was reported that this was a csi atherectomy procedure and atherectomy and stenting had been performed in the left anterior descending (lad) coronary artery with stenosis and calcification.The nc trek balloon dilatation catheter (bdc) was prepped and during advancement on the guide wire without resistance, the shaft became kinked before entering the patient.The device was used anyway and during advancement there was slight resistance with the anatomy and the shaft separated.The proximal portion was simply withdrawn but the distal portion was removed when it was trapped by an additional bdc.There were no adverse patient sequela.No additional information was provided.
 
Additional Manufacturer Narrative
The balloon dilatation catheter (bdc) was not returned for analysis.Based on the report information the account damaged the device during prep.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lot revealed no other complaints.The reported event stated the balloon was inflated to 20atms.It should be noted that instructions for use (ifu), "prior to use examine all equipment carefully for defects.Examine the dilatation catheter for bends, kinks, or other damage.Do not use any defective equipment.¿ it is likely the ifu violations contributed to the reported difficulties.Based on the information provided the cause for reported the difficulties was due to use error.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.D10, h3 - it was initially reported that the device would be returned for analysis.Subsequent information revealed that the device was retained by hospital and is not available for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15915753
Report Number2024168-2022-12222
Device Sequence Number15089096
Product Code LOX
UDI-Device Identifier08717648151880
UDI-Public08717648151880
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K103153
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2025
Device Model Number1012449-20
Device Catalogue Number1012449-20
Device Lot Number20928G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/14/2022
Supplement Date Received by Manufacturer12/22/2022
Initial Report FDA Received Date12/05/2022
Supplement Report FDA Received Date12/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/28/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNSPECIFIED GUIDE WIRE
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-