• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHT MEDICAL TECHNOLOGY INC PROPHECY INBONE PATIENT SPECIFIC GUIDES; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHT MEDICAL TECHNOLOGY INC PROPHECY INBONE PATIENT SPECIFIC GUIDES; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number PROPINB
Medical Device Problem Codes Inadequacy of Device Shape and/or Size (1583); Patient-Device Incompatibility (2682)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed. a review of the device history is not possible because the lot number was not communicated. should additional information become available, it will be provided on a supplemental report.
 
Event or Problem Description
It was reported that the prophecy blocks did not fit patient.The surgical team was 100% sure it was correct blocks for the patient.The blocks sat to medial.The physician had to cut the blocks to make them work.
 
Additional Manufacturer Narrative
Please note the correction made to the h6 device code: the reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.The prophecy team was made aware of the complaint and conducted an internal investigation into their processing of these guides.According to their review, "the ct scan was processed within normal, acceptable ranges no errors were identified during the segmentation procedure.The alignment guides were designed within normal, acceptable ranges no abnormalities in the internal process were found.The engineering drawings and quality documents were processed within normal, acceptable ranges.No abnormalities in the internal process were found." more detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
It was reported that the prophecy blocks did not fit patient.The surgical team was 100% sure it was correct blocks for the patient.The blocks sat to medial.The physician had to cut the blocks to make them work.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROPHECY INBONE PATIENT SPECIFIC GUIDES
Common Device Name
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15915799
Report Number3010667733-2022-00427
Device Sequence Number6798117
Product Code HSN
UDI-Device Identifier00889797057110
UDI-Public00889797057110
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K110360
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/06/2023
Device Model NumberPROPINB
Device Catalogue NumberPROPINB
Device Lot Number1744162
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/10/2022
Supplement Date Received by Manufacturer04/06/2023
Initial Report FDA Received Date12/05/2022
Supplement Report FDA Received Date05/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/06/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-