• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 NICORE NITINOL GUIDEWIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 NICORE NITINOL GUIDEWIRE Back to Search Results
Model Number 150NSA35
Medical Device Problem Codes Flaked (1246); Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/17/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the percutaneous nephrolithotomy was performed for the patient on (b)(6) at 13:00pm.Prior to use, the outer package was inspected for integrity and no abnormalities were found.Upon opening the outer package, the inner steel wire at the proximal end of the guide wire was found exposed externally.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
The reported event is confirmed, cause unknown.No physical sample was returned, however, a photo sample was submitted.Visual evaluation of the photo sample noted the core wire exposed at the tip of the guidewire.Although a specific cause cannot be determined, potential root causes for this event could be, "inadequate material selection, and/or bonding between layers, degradation".As no patient involvement was reported the device was not used, however it is intended for treatment purposes.It is unknown if the device had met all relevant specifications.No manufacturing issues or non-conformances were noted during review of the dhr that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "inspect all guidewires for damage prior to use.Bending or kinking during or prior to placement could damage the guidewire.Do not attempt to use the guidewire if it has been damaged.Use of a damaged wire may result in damage to the urinary tract." "do not reshape the guidewire by any means.Attempting to reshape the guidewire may cause damage resulting in release of fragments into the urinary tract.Failure to exercise proper caution may result in damage to the urinary tract." "do not manipulate or remove the guidewire through a metal cannula or needle.This may result in destruction/separation of the outer jacket of the wire requiring retrieval." "do not use dry gauze to manipulate the guidewire as this can damage the surface coating and make the wire tacky." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: the actual/suspected device was inspected.
 
Event or Problem Description
It was reported that the percutaneous nephrolithotomy was performed for the patient on (b)(6) at 13:00pm.Prior to use, the outer package was inspected for integrity and no abnormalities were found.Upon opening the outer package, the inner steel wire at the proximal end of the guide wire was found exposed externally.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NICORE NITINOL GUIDEWIRE
Common Device Name
GUIDEWIRE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key15915873
Report Number1018233-2022-09261
Device Sequence Number4702196
Product Code EZB
UDI-Device Identifier00801741051500
UDI-Public(01)00801741051500
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/16/2024
Device Model Number150NSA35
Device Catalogue Number150NSA35
Device Lot NumberNGET0766
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/18/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/18/2022
Supplement Date Received by Manufacturer04/04/2023
Initial Report FDA Received Date12/05/2022
Supplement Report FDA Received Date04/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/10/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-