| Catalog Number |
103401 |
| Medical Device Problem Codes |
Break (1069); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/05/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during priming with a cartridge dialyzer line, an external saline leak from the ¿normal saline line¿ was observed.It was further reported the heparin connection port is broken.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
The actual device was not received; however, eight (8) retained samples were evaluated.Visual inspection with the naked eye of the eight retention samples did not identify any abnormalities that could have contributed to the reported condition.Functional testing including leak testing was performed with no issues noted.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|