Model Number 130712000
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Osteolysis (2377)
Date of Event 01/01/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this was a rtsa performed in 2020.The surgery was completed successfully with no surgical delay.After surgery, on (b)(6) 2022 , the sales rep was informed by the surgeon that the patient had a delayed infection.The sales rep suggested to the surgeon that the treatment should be performed by debridement if the infection was mild or by removing the implant, doing debridement, and filling cement beads containing antibiotics if the infection was severe, etc.However, when the situation was checked again, the following situations were found.; some disease developed in the axilla after surgery, and the patient visited the dermatology department.Since the condition did not improve, the surgeon began to suspect that the infection was caused by the implants used in the initial surgery.Ulcer and granulation have formed in the axilla, and mrsa has been attached from the skin.The surgeon commented that the cause of infection was propionibacterium acnes.Slight osteolysis is observed on the stem site medially, but there is no instability both on the base plate site and stem site so far.Because the infection was expected to be quite severe, the sales rep suggested to the surgeon that prudent procedures be taken.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND MODULAR HUM STEM D12 HA
Common Device Name
DELTA XTEND IMPLANTS : SHOULDER HUMERAL STEMS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY INT'L LTD. 8010379
st anthonys road
leeds LS11 8 DT
UK   LS11 8 DT
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key15916300
Report Number1818910-2022-24322
Device Sequence Number6798133
Product Code KWS
UDI-Device Identifier10603295027171
UDI-Public10603295027171
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K192855
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number130712000
Device Catalogue Number130712000
Device Lot Number5348493
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/14/2022
Supplement Date Received by Manufacturer01/16/2023
Initial Report FDA Received Date12/05/2022
Supplement Report FDA Received Date01/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DXTEND GLENOSPHERE STD D38MM; DXTEND METAGLENE; DXTEND MOD EPI 1 ECC RIGHT HA; DXTEND MODULAR HUM STEM D12 HA; DXTEND SCREW LOCK D4.5X30MM; DXTEND SCREW LOCK D4.5X42MM; DXTEND SCREW NO LOCK D4.5X18MM; DXTEND STAND PE CUP D38 +3MM
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-