• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number TRLP-220
Medical Device Problem Codes Device Appears to Trigger Rejection (1524); Material Rupture (1546)
Health Effect - Clinical Codes Capsular Contracture (1761); Failure of Implant (1924)
Date of Event 11/02/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
 
Event or Problem Description
Healthcare professional reported right side rupture.Device remains implanted.
 
Event or Problem Description
Physician additionally reported "baker iii capsular contracture." the device has been explanted and replaced.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: b.5, d.6b, h.6.
 
Event or Problem Description
Healthcare professional reported right side hole-nonspecific via an rga form.
 
Additional Manufacturer Narrative
Device evaluation: based on the device analysis grid, the assessments of the complaint are: ¿ capsular contracture: unable to confirm as it is a medical event not related to the device.¿ rupture: observed broken sharp assessed as an unidentified (tear) opening.Additional observations: ¿ crease flat observed in the device.¿ red particles observed in the device.No further actions are required as the device was implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key15917782
Report Number9617229-2022-21517
Device Sequence Number6535862
Product Code FTR
UDI-Device Identifier10888628007857
UDI-Public10888628007857
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/22/2020
Device Model NumberTRLP-220
Device Catalogue NumberTRLP-220
Device Lot Number2857563
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/14/2022
Supplement Date Received by Manufacturer12/05/2022
01/12/2023
Initial Report FDA Received Date12/05/2022
Supplement Report FDA Received Date12/21/2022
01/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/22/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age27 YR
Patient SexFemale
Patient Weight67 KG
Patient EthnicityNon Hispanic
-
-