| Model Number |
TRLP-220 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Material Rupture (1546)
|
| Health Effect - Clinical Codes |
Capsular Contracture (1761); Failure of Implant (1924)
|
| Date of Event |
11/02/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Physician additionally reported "baker iii capsular contracture." the device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b.5, d.6b, h.6.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side hole-nonspecific via an rga form.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: ¿ capsular contracture: unable to confirm as it is a medical event not related to the device.¿ rupture: observed broken sharp assessed as an unidentified (tear) opening.Additional observations: ¿ crease flat observed in the device.¿ red particles observed in the device.No further actions are required as the device was implanted.
|
| |
|