MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC CATH PKGD: BERMAN 5 FR 80CM; CATHETER, FLOW DIRECTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC CATH PKGD: BERMAN 5 FR 80CM; CATHETER, FLOW DIRECTED Back to Search Results
Model Number IPN923529
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the catheter was found torn near the hub during the procedure.The catheter was replaced with a new one at the same insertion site.The user did not notice the tear at the beginning of the procedure.No harm to the patient was reported.Patient condition is reported as "fine".
 
Additional Manufacturer Narrative
(b)(4).
 
Additional Manufacturer Narrative
(b)(4).The reported lot number (16f21m0029) matches the lot number on the returned original packaging pouch/label.Returned for investigation was a 5fr.80cm berman catheter with the original packaging pouch.Upon return, the catheter body was immediately noted ruptured near the junction hub; the body was noted ruptured from approximately 88.9cm to 89.5cm from the distal tip of the catheter.The supplied control stroke syringe was noted connected to the inflation lumen stopcock; no damage or abnormalities were noted to the returned syringe.The inflation lumen stopcock was in the closed position.The recommended volume capacity of the balloon is 0.75cc.Upon microscopic inspection, a wrinkle was noted on the balloon; no damage was noted to the balloon.No condensation was noted within the inflation lumen extension line.No contrast media was noted within the injection lumen extension line.Dried blood/contrast media was noted within the berman holes (injection lumen) near the catheter tip.Dried blood was noted on the exterior of the sample.No other damage or abnormalities were noted.The inflation lumen was injected with 0.75cc of air using the returned control stroke syringe.The balloon inflated symmetrically.One side of the balloon measured approximately 4mm.The other side measured approximately 4mm.The inflation lumen was injected with 0.75cc of air using the returned control stroke syringe.The balloon inflated symmetrically.The balloon deflated in less than 3 seconds when the syringe was removed.No pull away was noted after the tug test.The balloon was placed in water and air was injected into the inflation lumen again.No leak was noted.The damaged/ruptured catheter body did not result in damage to the inflation lumen.The catheter's injection lumen was aspirated/flushed, and air was noted leaking from the damaged/ruptured catheter body.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.The reported complaint of catheter body cut/torn during use is confirmed.The catheter body was found ruptured near the junction hub upon receipt of the sample.Based on a review of the device history record (dhr), the product met specification upon release; however, the specifications were not met during the complaint investigation due to the ruptured catheter body.The root cause of the ruptured catheter body is undetermined.Other remarks: n/a.Corrected data: n/a.
 
Event or Problem Description
It was reported that the catheter was found torn near the hub during the procedure.The catheter was replaced with a new one at the same insertion site.The user did not notice the tear at the beginning of the procedure.No harm to the patient was reported.Patient condition is reported as "fine".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CATH PKGD: BERMAN 5 FR 80CM
Common Device Name
CATHETER, FLOW DIRECTED
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
bryanna connelly
3015 carrington mill blvd
morrisville 27560
MDR Report Key15919576
Report Number3010532612-2022-00548
Device Sequence Number4459835
Product Code DYG
UDI-Device Identifier10801902206654
UDI-Public10801902206654
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K892530
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2023
Device Model NumberIPN923529
Device Catalogue NumberAI-07135
Device Lot Number16F21M0029
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/01/2022
Supplement Date Received by Manufacturer01/20/2023
Initial Report FDA Received Date12/05/2022
Supplement Report FDA Received Date01/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-