• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON BD MICROTAINER® CONTACT-ACTIVATED LANCET; MANUAL SURGICAL INSTRUMENT FOR GENERAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON BD MICROTAINER® CONTACT-ACTIVATED LANCET; MANUAL SURGICAL INSTRUMENT FOR GENERAL USE Back to Search Results
Catalog Number 366592
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code Needle Stick/Puncture (2462)
Date of Event 11/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation summary: bd had not received samples or photos for investigation.Therefore, retention samples from bd inventory were evaluated by functional testing and no issues were observed relating to retraction as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure modes retraction/needle stick injury.Bd was not able to identify a root cause for the indicated failure modes.Oem manufacturer: the manufacturing location for this product is (b)(4).This site is an oem manufacturing site.(b)(4).
 
Event or Problem Description
It was reported when using the bd microtainer® contact-activated lancet the nurse stuck her finger with the blade of lancet.The following information was provided by the initial reporter.The customer stated: according to the customer's report, the nurse stuck her finger with the blade of lancet.Whether the blade was not retracted or whether this is attributed to the nurse's operation is unknown.
 
Additional Manufacturer Narrative
H.10.Corrected information: oem manufacturer: the manufacturing location for this product is htl.
 
Event or Problem Description
It was reported when using the bd microtainer® contact-activated lancet the nurse stuck her finger with the blade of lancet.The following information was provided by the initial reporter.The customer stated: according to the customer's report, the nurse stuck her finger with the blade of lancet.Whether the blade was not retracted or whether this is attributed to the nurse's operation is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MICROTAINER® CONTACT-ACTIVATED LANCET
Common Device Name
MANUAL SURGICAL INSTRUMENT FOR GENERAL USE
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15920630
Report Number2243072-2022-02095
Device Sequence Number11861273
Product Code FMK
UDI-Device Identifier50382903665926
UDI-Public50382903665926
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number366592
Device Lot NumberB9H38H6
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/15/2022
Supplement Date Received by Manufacturer11/15/2022
Initial Report FDA Received Date12/05/2022
Supplement Report FDA Received Date12/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;