| Model Number |
I400+ |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of questionable gen.2 ise indirect for na results for 1 patient sample on a cobas integra 400 plus module.The initial na result was 105 mmol/l.The sample was repeated on another integra module and the result was 120 mmol/l.The initial result was not reported outside of the laboratory.The na electrode lot number and expiration date were requested but not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer replaced the sodium electrode.The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation did not identify a product problem.The root cause of the event could not be determined.
|
| |
|