| Model Number |
DFR00V |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Eye Injury (1845)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that an intraocular lens (iol) was explanted from the patient's operative eye for an unspecified reason.Another johnson & johnson lens, model dib00 21.0 diopter was implanted as a replacement.No other information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Age or date of birth, weight, ethnicity: information unknown/not provided.Date of event: unknown/not provided.If implanted; give date: unknown/not provided.The device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain missing information; however, no definitive response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional info: device available for evaluation? yes.Returned to manufacturer on: jan 18, 2023.Device evaluated by manufacturer¿ yes.Device evaluation: the suspect iol was received wrapped in gauze inside of a plastic bag and a samples return form were received.The iol was received cut into multiple pieces; therefore, further evaluation could not be performed.The complaint issue were not confirmed.The other observed issue during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Conclusion: as a result of the investigation there is no indication of a product quality deficiency, and the reported issue could not be confirmed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|