• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CORPORATION, ASHITAKA CAPIOX HOLLOW FIBER OXYGENATOR/ARTERIAL FILTER; OXYGENATOR, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CORPORATION, ASHITAKA CAPIOX HOLLOW FIBER OXYGENATOR/ARTERIAL FILTER; OXYGENATOR, CARDIOPULMONARY BYPASS Back to Search Results
Model Number N/A
Medical Device Problem Codes Break (1069); Increase in Pressure (1491)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/26/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The user facility reported that when the caipox device oxygenation, was high pressured.Overpressure of the oxygenator was hypothesized with a possible breakage of oxygenating fiber that has determined gaseous embolism along the arterial line of the cec circuit.There were no consequences and was a regular post operative course.The problem occurred during a cardio bypass before x-clamps.The event resulted in a 20-minute delay.The product was changed out, capiox fx05.No consequences for the patient.As a result of the reported issue 380ml of blood was lost.Flow was 650ml/minutes and pressure out were 600mmhg.The event occurred intra-operative.The device malfunction due to increasing pressure.The procedure outcome was completed successfully.The final patient impact was not harmed.
 
Additional Manufacturer Narrative
Patient ethnicity: requested, not provided.Patient race: requested, not provided.Implanted date: device was not implanted.Explanted date: device was not explanted.Initial reporter health professional: requested, unknown.Initial reporter occupation: vigilance responsible.Pma/510(k): k071572, k130280.The actual device has not been returned for evaluation.The investigation is currently ongoing.A follow-up report will be submitted once the investigation is complete.Review of the manufacturing record and shipping inspection record of the actual sample found no anomaly in them.A search of the complaint file found no similar indications regarding the involved product code/lot number combination.Terumo medical products (tmp) (importer) registration no.(b)(4) is submitting this report on behalf of ashitaka factory of terumo corporation (manufacturer) registration no.(b)(4).
 
Additional Manufacturer Narrative
This report is being submitted as follow-up no.1 to provide the device return date in section d9, to update section h3, and to provide the completed investigation results.The actual device has been returned for evaluation.Visual inspection of the actual sample upon receipt revealed that no anomaly such as a brakeage was found.After rinsing and drying the actual sample, the oxygenator was installed into a circuit consisting of tubes, and bovine blood (hct: 35%, temp.: 37°c) was circulated with a flow rate of 1.5 l/min.Then, the pressure drop was measured.It was confirmed to meet the factory's specifications and no obstruction was found.After circulation of bovine blood, saline solution was flowed in the blood channel.No blood clot that led to the obstruction was formed.Review of the manufacturing record and the shipping inspection record of the actual sample confirmed that there was not any anomaly in them.A search of the complaint file found no other similar report of the product with the involved product code/lot number combination.Based on the investigation result, no anomaly was found in the actual sample after rinsing.As a cause of occurrence, it was likely that blood-derived obstruction occurred for some reason.However, since no anomaly was found in the performance test of actual sample, it was not possible to clarify the cause of this complaint from the investigation result.Relevant instructions for use (ifu) reference: the end of the report do not reduce heparin during circulation.Otherwise, blood clotting might occur.(warnings) adequate heparinization of the blood is required to prevent it from clotting in the system.(warnings).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPIOX HOLLOW FIBER OXYGENATOR/ARTERIAL FILTER
Common Device Name
OXYGENATOR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
TERUMO CORPORATION, ASHITAKA
150 maimaigi-cho
fujinomiya city, 418
JA  418
Manufacturer (Section G)
TERUMO CORPORATION, ASHITAKA
reg. no. 9681834
150 maimaigi-cho
fujinomiya city, 418
JA   418
Manufacturer Contact
gina digioia
reg. no. 2243441
950 elkton blvd.
elkton, MD 21921
6402040886
MDR Report Key15921016
Report Number9681834-2022-00241
Device Sequence Number20173742
Product Code DTZ
UDI-Device Identifier04987350781772
UDI-Public04987350781772
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K071572
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2024
Device Model NumberN/A
Device Catalogue NumberCX*FX05RW
Device Lot Number211220
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/08/2022
Supplement Date Received by Manufacturer01/11/2023
Initial Report FDA Received Date12/05/2022
Supplement Report FDA Received Date01/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age2 YR
Patient SexFemale
Patient Weight9 KG
-
-