BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA. BD ANGIOCATH¿ IV CATHETER; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 381112
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medical device expiration date: unknown.Initial reporter phone#: unknown.Initial reporter e-mail: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Event or Problem Description
It was reported that the bd angiocath¿ iv catheter leaked during use.The following information was provided by the initial reporter, translated from japanese: "this is a report about leakage occurring during the use of iagbc.During infusion with the use of iagbc, leakage occurred with leakage site unknown.".
 
Additional Manufacturer Narrative
H6: investigation summary: bd received a 24 gauge angiocath device from an unknown lot for evaluation.A review of the device history record could not be performed as the reported lot was unknown.However, due to difficulties returning samples to the manufacturing facility our quality engineer had to investigate the returned sample using high resolution photographs of the defect.The engineer was able to identify that the catheter was crumpled but there appeared to be no damage to the rest of the unit.Leakage testing was not able to be performed due to limitations and the site the sample was returned to.Therefore, based off the provided photos the engineer was unable to verify the reported defect of leakage or determine a definitive root cause.
 
Event or Problem Description
It was reported that the bd angiocath¿ iv catheter leaked during use.The following information was provided by the initial reporter, translated from japanese: "this is a report about leakage occurring during the use of iagbc.During infusion with the use of iagbc, leakage occurred with leakage site unknown.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ANGIOCATH¿ IV CATHETER
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA.
ave. pres.juscelino kubitschek
273 francisco bernardino
juiz de fora
Manufacturer (Section G)
BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA.
ave. pres.juscelino kubitschek
273 francisco bernardino
juiz de fora
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15921239
Report Number9610048-2022-00162
Device Sequence Number2492187
Product Code FOZ
UDI-Device Identifier00382903811120
UDI-Public(01)00382903811120
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K151698
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number381112
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/15/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/09/2022
Supplement Date Received by Manufacturer03/09/2023
Initial Report FDA Received Date12/05/2022
Supplement Report FDA Received Date03/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-