• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENDOLOGIX AFX2; BIFURCATED STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENDOLOGIX AFX2; BIFURCATED STENT GRAFT Back to Search Results
Model Number BEA28-100/I16-40
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Codes Aneurysm (1708); Failure of Implant (1924); Clinical Endoleak (4843)
Date of Event 11/16/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The patient was initially treated for an abdominal aortic aneurysm (aaa) with the implant of an afx2 bifurcated stent graft.Approximately three (3) years post-initial procedure, during a routine follow-up, the patient was identified with an endoleak type 3b and an aneurysm enlargement.The patient is scheduled for a re-intervention.
 
Additional Manufacturer Narrative
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2.Device remains implanted.
 
Event or Problem Description
The patient was initially treated for an abdominal aortic aneurysm (aaa) with the implant of an afx2 bifurcated stent graft.Approximately three (3) years post-initial procedure, during a routine follow-up, the patient was identified with an endoleak type 3b and an aneurysm enlargement.The patient is scheduled for a re-intervention.Additional information: a re-intervention was done.The physician elected to implant an alto main body, two (2) ovation ix iliac limbs and ovation prime fill polymer.The final patient status was reported as doing well and discharged to go home.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good-faith efforts to retrieve a reported adverse event/incident-related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the type iiib endoleak and aneurysm enlargement are unconfirmed.The additional endovascular procedure complaint is confirmed.This is moderately consistent with the reported adverse event/incident.It was reported that at index "placed only a mainbody graft and it lost seal in the neck, so it was a type 1a" (versus a type iiib endoleak), however, this could not be independently verified due to lack of imaging.Procedure-related harms, device, user, procedure, or anatomy-relatedness of this complaint could not be determined with the medical records available for review.The additional endovascular procedure was confirmed.The final patient status was reported as doing well and discharged to go home.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx2.Corrections: b2: outcomes attributed to adverse events has been updated b5 describe event or problem h6 health effect - clinical code; remove 1924 h6 investigation finding codes; remove 3233 h6 investigation conclusion codes; remove 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX2
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer (Section G)
ENDOLOGIX
2 musick
,
irvine CA 92618
Manufacturer Contact
gary kirchgater
2 musick
,
irvine, CA 92618
8009832284
MDR Report Key15921306
Report Number2031527-2022-00272
Device Sequence Number5446131
Product Code MIH
UDI-Device Identifier00818009014450
UDI-Public(01)00818009014450(17)180825
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall,Other
Type of Report Initial,Followup
Report Date (Section B) 11/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/25/2020
Device Model NumberBEA28-100/I16-40
Device Lot Number1732114-002
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/16/2022
Supplement Date Received by Manufacturer02/23/2023
Initial Report FDA Received Date12/05/2022
Supplement Report FDA Received Date03/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/14/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0007-2019
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age80 YR
Patient SexMale
-
-